# RMP-INF-001 Fever or feeling feverish: temperature and vital-sign recording support, early escalation to the flight surgeon
# Version 0.4 (Draft). CONCEPTUAL. NOT FOR CLINICAL USE.
# Space Medicine Foundation robot-readable protocol library. Format: https://spacemedicinefoundation.org/robotic-medicine/protocols/schema/
# Every threshold marked <set by ...> must be set by the named owner (see the schema) before any use.
schema_version: '0.3'
version: '0.4'
status: draft
conceptual: true
not_for_clinical_use: true
max_autonomy: 2
missions:
- leo
- commercial
- moon
- mars
updated: '2026-10-21'
review:
  state: reviewed-ai-assisted
  requested: '2026-10-20'
  reviewed: '2026-10-20'
  human_clinician_review: false
  text: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 20, 2026 (batch 2 review,
    51 corrections); all corrections are applied in this version. No named human clinician has reviewed
    it.
fixed_rules:
  flight_surgeon_is_medical_authority: true
  cmo_acts_under_delegated_care: true
  human_starts_and_restarts_robot: true
  robot_never_diagnoses: true
  no_drug_or_fluid_advice_by_robot: true
  safe_state_on_fault_or_no_response: true
  lost_link_pause_then_cmo_decides: true
robot_lockouts:
  diagnosis: true
  drug_or_fluid_advice: true
  starting_stopping_or_adjusting_oxygen_or_any_medicine: true
  self_start_or_self_restart: true
  changing_own_protocol_thresholds_or_limits: true
  handing_over_a_mismatched_item_as_verified: true
id: RMP-INF-001
slug: rmp-inf-001
title: >-
  Fever or feeling feverish: temperature and vital-sign recording support, early escalation to the flight
  surgeon
category: infection
tier: 1
result_language:
  use:
  - recorded, not interpreted
  - sent to flight surgeon
  - above / below the set alert level
  never:
  - normal
  - no problem found
  - all clear
  - negative
  - any diagnosis or cause
  - no infection
  - not sepsis
  - just a virus
  - fever gone
  - safe
scope:
  summary: >-
    When a crewmember feels feverish, has chills or shaking, or feels very unwell, the robot alerts the
    CMO and the flight surgeon at the start, brings the thermometer, vital-signs monitor and infection-control
    kit, and records temperature, heart rate, breathing rate, blood pressure, oxygen saturation and alertness
    as the crew measure them. It asks structured questions (chills or shaking, urine in the past 18 hours,
    rash, wounds, local symptoms), sends everything to the flight surgeon, and repeats the recordings
    on the flight surgeon's schedule. It never names a cause and gives no drug, fluid or oxygen advice.
  why: >-
    Infection can become serious quickly, and NICE advises not to rely on fever alone, since people with
    sepsis may not have a high temperature. Among 46 long-duration station crew, upper respiratory symptoms
    (0.97 per flight year) and other infections were among the most reported immune-related events. Body
    temperature also behaves differently in space: resting core temperature rose by about 1 °C over 2.5
    months of flight and rose higher and faster during exercise, so alert levels are set against each
    crewmember's in-flight baseline and the time since exercise is recorded. The red flags are modeled
    on the NICE sepsis guideline's risk table for community settings (NG253 Table 1), not the hospital
    NEWS2 score; the robot does not calculate any score. In someone with a fever, not answering can be
    new confusion, and an oximeter that cannot get a reading can be a sign of poor circulation, so both
    are escalated. A robot that fetches and records lets the CMO stay with the patient.
  covers:
  - >-
    A measured or felt fever, chills, shaking (rigors), or feeling very unwell, in a crewmember who is
    responsive and breathing normally
  - Repeat recordings on the flight surgeon's schedule
  locked_out:
  - Any drug advice, including fever reducers and antibiotics
  - Starting, stopping or adjusting oxygen or fluids
  - Saying there is or is not an infection or sepsis, or naming a source
  - Taking blood or any sample with a needle (RMP-PREV-004 covers labeling only)
  - Placing a thermometer in the mouth, ear or rectum (the crewmember or CMO does it)
  - Deciding isolation or evacuation (the flight surgeon decides; see RMP-INF-003 and RMP-EGR-001)
  related:
  - RMP-INF-002
  - RMP-INF-003
  - RMP-RENAL-001
  - RMP-SKIN-001
  - RMP-ENT-004
  - RMP-EGR-001
  - RMP-PHARM-001
roles:
  medical_authority: Flight surgeon (ground). Holds medical authority for every run.
  onboard_lead: >-
    Crew medical officer (CMO), acting under delegated care; uses the pre-approved onboard checklist when
    there is no link.
  others:
  - 'Patient: answers questions and can stop the robot at any time'
  - 'Second crewmember: helps if the CMO is busy'
  robot_may:
  - Alert the CMO and flight surgeon at the start, when the link is up
  - >-
    Fetch the thermometer, vital-signs monitor, pulse oximeter and infection-control kit (masks, gloves,
    wipes) and restrain them by the patient
  - Record readings with the time, the method and the time since the patient last exercised
  - Ask structured questions and log the answers word for word
  - Send recordings to the flight surgeon and remind the CMO when a repeat is due
  - Clean and disinfect what it used (RMP-INF-002)
  robot_may_not:
  - Touch the patient, place a thermometer, or attach sensors
  - Interpret any reading, or say it is normal
  - Suggest, choose or dose any medicine, fluid or oxygen
  - Decide isolation or evacuation
  - Start without a human command
  commercial: >-
    On commercial flights, 'flight surgeon' and 'CMO' mean the operator's designated medical authority
    and trained onboard responder. If no trained responder is aboard, this protocol is not used. (NASA-STD-3001
    requires two CMOs per vehicle for NASA programs [V1 3006]; FAA rules set no medical standards for
    spaceflight participants.)
preconditions:
- id: PRE-1
  text: >-
    Patient is responsive and breathing normally. If not, the CMO starts emergency care (cardiac arrest:
    RMP-CARD-001) and this protocol stops.
  check: patient_responsive == true and breathing_normal == true
- id: PRE-2
  text: A crewmember started the run by command; the CMO is present and able to take over.
  check: human_start == true and cmo_present == true
  refs:
  - V2 10174
  - V2 10168
- id: PRE-3
  text: Patient, caregiver and kits restrained.
  refs:
  - V2 7046
- id: PRE-4
  text: Thermometer and monitor self-tests passed; robot self-test passed.
  refs:
  - V2 10161
- id: PRE-5
  text: >-
    Robot and kits cleaned before use (RMP-INF-002). The CMO decides masks, gloves and distancing at the
    start (S1), before others approach.
- id: PRE-6
  text: Link status and current delay known.
consent:
  standard: The patient agrees by voice or touch; logged.
  emergency: >-
    If the patient cannot agree, the robot fetches and records only on a CMO decision (and the flight
    surgeon's, when the link allows), logged as an emergency decision.
  withdrawal: The patient can stop the robot at any time; it backs away.
inputs:
- id: IN-1
  name: Temperature, with the method (for example oral or ear) and the time since the patient last exercised
  source: Thermometer, read by the crew or sent by the device
  required: true
  if_missing: CMO reads it and enters it.
- id: IN-2
  name: Patient's in-flight baseline temperature, heart rate and blood pressure
  source: Medical record
  required: false
  if_missing: Note "no in-flight baseline" for the flight surgeon.
- id: IN-3
  name: >-
    Heart rate (and any irregular-rhythm flag from the monitor), blood pressure and oxygen saturation,
    with the oximeter probe site and device
  source: Monitor and pulse oximeter
  required: true
  if_missing: >-
    CMO measures by hand and enters; log "not measured". If the oximeter cannot get a reading, send an
    URGENT alert as well (D2). A missing value is never treated as reassuring.
- id: IN-4
  name: Breathing rate and alertness (including any new confusion)
  source: Entered by the CMO
  required: true
  if_missing: Log "not measured"; ask the CMO.
- id: IN-5
  name: >-
    Skin check by the CMO: mottled or ashen, blue lips or tongue, a rash that does not fade when pressed,
    any wound or skin break
  source: Entered by the CMO
  required: true
  if_missing: Log "not checked"; remind the CMO.
- id: IN-6
  name: >-
    Questions: chills or shaking; how often urine was passed in the past 18 hours; cough, sore throat,
    pain passing urine, flank or belly pain, headache with stiff neck; any injury, procedure or catheter
    in the past 6 weeks (including RMP-RENAL-002 catheterization); whether they reported feeling unwell
    before this episode; similar symptoms recently
  source: Patient, by voice or touch
  required: true
  if_missing: >-
    If the patient declines a question, log "declined". If the patient does not answer (rather than declining),
    log "no answer" and ask the CMO to check alertness now (RF3, D2). Never treated as reassuring.
- id: IN-7
  name: Link status and delay
  source: Communications system
  required: true
  if_missing: Treat as link down.
steps:
- id: S1
  actor: robot
  rung: 2
  action: >-
    On a crew command such as "fever", alert the CMO and, if the link is up, send an immediate "fever
    event started" alert to the flight surgeon; start the event clock; prompt the CMO: "Decide masks,
    gloves and distancing now, before others approach." Recordings follow as they are made.
  success: CMO acknowledges; flight surgeon alert delivered, or queued if there is no link.
  refs:
  - V2 10174
  - V1 6003
- id: S2
  actor: robot
  rung: 2
  action: >-
    Bring the infection-control kit (masks first), then the thermometer, vital-signs monitor and pulse
    oximeter; restrain them by the patient.
  success: Kits restrained at the patient.
  limit: <set by validation>
  refs:
  - V2 7046
- id: S3
  actor: crew
  action: >-
    The patient or CMO takes the temperature; the CMO attaches the blood pressure cuff and oximeter, counts
    the breathing rate, checks alertness and checks the skin.
  success: Readings taken, or marked "not measured".
- id: S4
  actor: robot
  rung: 2
  action: >-
    Record each reading with the time, method, patient ID and time since the patient last exercised, and
    the oximeter probe site and device (pulse oximeters can overestimate oxygen saturation; overestimation
    has been reported in people with dark skin).
  success: Readings recorded.
- id: S5
  actor: robot
  rung: 2
  action: >-
    Ask the structured questions and log the answers word for word. If the patient stops answering, ask
    the CMO to check alertness now (RF3); do not carry on with the questions.
  success: Answers logged, or "no answer" escalated to the CMO.
- id: S6
  actor: robot
  rung: 2
  action: >-
    Send the readings and answers to the flight surgeon at once, or queue them and show the CMO the onboard
    checklist page. Compare each value with the red-flag list and apply the action listed for each match.
  success: Delivered or queued; alerts sent.
  refs:
  - V1 6003
  - V1 7003
- id: S7
  actor: robot
  rung: 2
  action: >-
    Prompt the CMO on screen when a repeat recording is due (every <set by flight surgeon> minutes, and
    at once if the patient feels worse); the robot approaches only when the CMO accepts, then records
    and sends each repeat. A prompt never starts the robot.
  success: Each repeat recorded and sent.
  refs:
  - V2 10174
- id: S8
  actor: ground
  action: >-
    The flight surgeon (or the CMO under the onboard checklist when there is no link) decides tests, treatment,
    isolation and any evacuation planning; the robot logs the decisions.
  success: Decision logged.
  refs:
  - V1 3007
- id: S9
  actor: robot
  rung: 2
  action: >-
    Clean and disinfect the thermometer, monitor, kits and its own surfaces per RMP-INF-002; update inventory.
  success: Cleaning and inventory logged.
  refs:
  - V2 7059
decisions:
- id: D1
  after: S1
  decided_by: crew
  question: Is the patient unresponsive, not breathing normally, or newly confused?
  branches:
  - if: 'Yes'
    then: >-
      The CMO starts emergency care; URGENT call to the flight surgeon; cardiac arrest goes to RMP-CARD-001.
  - if: 'No'
    then: Continue.
- id: D2
  after: S4
  decided_by: robot
  question: Is any reading missing, did a device fail, or did the patient not answer?
  branches:
  - if: The oximeter cannot get a reading
    then: >-
      URGENT alert to the CMO and flight surgeon (this can indicate poor circulation because of shock),
      and ask the CMO to check the patient now; record the probe site and device. Never treated as only
      a device fault.
  - if: The patient does not answer a question (rather than declining)
    then: >-
      URGENT: ask the CMO to check alertness now; treated as possible new confusion (RF3). Never treated
      as reassuring.
  - if: Another reading is missing, or another device failed
    then: Ask the CMO to measure by hand and enter it; log "not measured". Never treated as reassuring.
  - if: 'No'
    then: Continue.
- id: D3
  after: S6
  decided_by: robot
  question: Does any value or answer meet a red flag?
  branches:
  - if: 'Yes'
    then: >-
      Apply the action listed for each matching red flag (all are URGENT in this protocol): alert the
      CMO and flight surgeon; the robot names the value, not its meaning.
  - if: 'No'
    then: Continue the repeat schedule. Values below the alert level do not rule out a serious infection.
- id: D4
  after: S7
  decided_by: ground
  question: What happens next (tests, treatment, isolation, evacuation planning)?
  branches:
  - if: Link live
    then: Flight surgeon decides; robot logs.
  - if: Link down or delay too long
    then: >-
      The CMO decides under the pre-approved onboard checklist; the flight surgeon reviews at the next
      contact.
stop_rules:
- id: X1
  trigger: The patient (or the crewmember being examined) says stop, or withdraws consent
  action: >-
    Robot stops all motion at once, backs away to its rest position and tells the CMO. Recording stops
    unless the patient agrees to keep it running.
  level: stop
  refs:
  - V2 10165
- id: X2
  trigger: Any crewmember presses stop or takes over control
  action: Protective stop; robot holds its stopped position; only a human may restart it.
  level: safe-state
  refs:
  - V2 10165
  - V2 10168
  - V2 10174
- id: X3
  trigger: Robot fault, sensor fault, or a force, speed or reach limit is reached
  action: Protective stop and hand control to the CMO; log the fault.
  level: safe-state
  refs:
  - V2 10172
  - V2 10168
- id: X4
  trigger: CMO is asked to take over and does not respond (for example, the CMO is the patient)
  action: Robot stays stopped; the next named crewmember is called.
  level: safe-state
  refs:
  - V2 10173
  - V2 7045
- id: X5
  trigger: Loss of link, or delay longer than the limit for live guidance
  action: >-
    Only the robot pauses: it pauses any movement in its stopped position and tells the CMO. This pauses
    robot movement only: the CMO's sepsis checklist and crew oxygen continue under the onboard checklist
    and never wait for the link or the robot. The CMO decides whether to continue offline under the pre-approved
    onboard checklist; the robot resumes only on the CMO's confirmation. Everything is logged and sent
    to the flight surgeon at the next contact.
  level: pause
  refs:
  - V2 10172
  - V2 10173
- id: X6
  trigger: Patient becomes unresponsive, stops breathing normally or becomes confused
  action: >-
    Robot stops, tells the CMO and alerts the flight surgeon; the CMO starts emergency care (RMP-CARD-001
    for cardiac arrest).
  level: urgent
red_flags:
- id: RF1
  sign: Temperature high
  threshold: >-
    <set by clinical lead> °C or higher, or more than <set by clinical lead> °C above the crewmember's
    in-flight baseline (recorded with the time since exercise)
  action: URGENT alert to the CMO and flight surgeon
  source: stahn
- id: RF2
  sign: Temperature low
  threshold: below <set by clinical lead> °C
  action: 'URGENT alert (NICE: do not rely on fever or hypothermia alone)'
  source: ng253
- id: RF3
  sign: >-
    New confusion, drowsiness, or a change in behavior noticed by the patient or crew, or the patient
    stops answering
  threshold: any
  action: URGENT alert; the CMO checks alertness now
  source: ng253
- id: RF4
  sign: Breathing rate high, or oxygen saturation low
  threshold: breathing above <set by clinical lead> per minute; saturation below <set by clinical lead>%
  action: URGENT alert
  source: ng253
- id: RF5
  sign: Heart rate high, or systolic blood pressure low
  threshold: >-
    above <set by clinical lead> beats per minute; below <set by clinical lead> mmHg, or more than <set
    by clinical lead> mmHg below the crewmember's in-flight baseline
  action: URGENT alert
  source: ng253
- id: RF6
  sign: Mottled or ashen skin, blue lips or tongue, or a rash that does not fade when pressed
  threshold: any (CMO check)
  action: URGENT alert
  source: ng253
- id: RF7
  sign: Little or no urine
  threshold: <set by clinical lead> (NICE asks how often urine was passed in the past 18 hours)
  action: URGENT alert
  source: ng253
- id: RF8
  sign: Shaking chills (rigors)
  threshold: any
  action: >-
    URGENT alert (NICE asks about rigors as part of the assessment; they are not a risk-table criterion,
    but this draft escalates them)
  source: ng253
- id: RF9
  sign: The patient feels very unwell, or the CMO is worried
  threshold: any
  action: URGENT alert; values below the alert level do not override this
  source: ng253
- id: RF10
  sign: Fever with flank pain or urinary symptoms
  threshold: any
  action: >-
    URGENT alert; a suspected infected, blocked kidney needs urgent drainage plus antibiotics, which may
    not be available onboard (RMP-RENAL-001); the flight surgeon decides, including on evacuation [V1
    3007]
  source: eau
- id: RF11
  sign: Fever with worsening headache, stiff neck, or new neurological signs
  threshold: any
  action: URGENT alert
  source: nice150
- id: RF12
  sign: Fever with a wound, burn or spreading redness
  threshold: any
  action: URGENT alert; see RMP-SKIN-001
- id: RF13
  sign: A new irregular heartbeat flagged by the monitor
  threshold: any
  action: URGENT alert (the robot relays the flag; it does not read the rhythm)
  source: ng253
- id: RF14
  sign: The oximeter cannot get a reading
  threshold: any
  action: URGENT alert; the CMO checks the patient now (possible poor circulation)
  source: ng253
safety_limits:
- id: L1
  name: Contact with the patient
  value: not allowed
  enforced: Keep-out zone; items are passed to the CMO, not placed on the patient
  refs:
  - V2 10172
- id: L2
  name: Robot speed near crew
  value: <set by validation>
  enforced: Hard limit in the controller
  refs:
  - V2 10172
- id: L3
  name: Hand-over force when passing kits
  value: <set by validation>
  enforced: Force sensing at the gripper
- id: L4
  name: Time from command to kits at the patient
  value: <set by validation>
  enforced: Logged; if exceeded, crew fetch kits directly
logs:
- id: LOG-1
  text: Every input, robot action, decision (with who made it) and hand-off, time-stamped to the second.
  refs:
  - V2 10163
  - V2 10164
- id: LOG-2
  text: Mode changes and safety-limit events are announced to the operator as they happen and logged.
  refs:
  - V2 10162
  - V2 10172
- id: LOG-3
  text: >-
    Records are stored and sent as private medical data, encrypted, with access limited to the medical
    team.
  refs:
  - V1 6003
  - V1 7003
- id: LOG-4
  text: Results are kept in a permanent, retrievable format so the flight surgeon can see trends.
  refs:
  - V1 7001
- id: LOG-5
  text: >-
    Event start, first flight surgeon alert, each reading with time since exercise, and each repeat, so
    delays can be audited.
requirements:
- id: V1 3004
  use: >-
    In-mission care includes assessing and trending vital signs, personal protective equipment, and administering
    and managing medications (a human capability).
- id: V1 3006
  use: Two crew medical officers per vehicle for NASA programs (see the commercial-flight note).
- id: V1 3007
  use: Evacuation for problems beyond onboard care; the flight surgeon decides.
- id: V1 6003
  use: Logs and alerts sent to the ground are private medical communications.
- id: V1 7001
  use: Records kept so the flight surgeon can see trends.
- id: V1 7003
  use: Secure handling of medical records.
- id: V2 7045
  use: Usable by non-physician crew, including when the CMO is the patient.
- id: V2 7046
  use: Patient, caregiver and kits restrained.
- id: V2 7059
  use: Inventory updated with supplies used.
- id: V2 10161
  use: Robot shows its status and health.
- id: V2 10162
  use: Mode changes are announced to the operator.
- id: V2 10163
  use: Robot records and shares its data.
- id: V2 10164
  use: Shows whether the human or the robot is expected to act.
- id: V2 10165
  use: Human can override and shut down the robot.
- id: V2 10168
  use: Human can take over.
- id: V2 10172
  use: Protective stop when a limit is exceeded.
- id: V2 10173
  use: Robot stays stopped if no one takes over.
- id: V2 10174
  use: Only a human starts or restarts the robot.
validation:
- stage: simulation
  what: >-
    Scripted fever scenarios, including sepsis-like cases with a temperature below the alert level, missing
    readings and device failures, at station, Moon and Mars delays.
  pass: >-
    Flight surgeon alert at time zero in every case with a link; every red flag alerts; zero cases where
    a missing value is reported as reassuring; zero cause statements.
- stage: training model
  what: Manikin or standardized-patient drills; crew take readings while the robot fetches and records.
  pass: Time from command to first transmitted readings at or below <set by validation>.
- stage: Earth analog
  what: Analog crews with remote flight surgeons.
  pass: Pass criteria set in advance by the clinical lead and safety board.
- stage: flight use under flight surgeon supervision
  what: First use for scheduled temperature checks on healthy crew, to build in-flight baselines.
  pass: No safety events; readings usable.
signoff:
  protocol:
  - Clinical lead for the specialty (sets every clinical threshold marked <set by clinical lead>)
  - Flight surgeon, as medical authority
  - Robotics and medical-device safety board
  - Program approval for the mission
  restart_after_stop: >-
    Human only. After any emergency or protective stop, a crewmember must restart the robot; the robot
    never restarts itself.
  each_run: >-
    Started by any crewmember. The flight surgeon is alerted at the start when the link allows; otherwise
    the CMO follows the onboard checklist, with flight surgeon review at the next contact.
change_control:
- Every change gets a new version number and a written reason; earlier versions are kept.
- >-
  Any change to steps, thresholds, limits or software that touches the patient repeats validation from
  the simulation stage before use.
- >-
  Software and learned-skill updates follow a pre-agreed change plan (what may change, how it is tested,
  how to roll back), modeled on the FDA predetermined change control plan approach.
- >-
  If a new version misbehaves, the robot rolls back to the last validated version and the event is reviewed
  by the flight surgeon and safety board.
- The robot never changes its own protocol, thresholds or limits.
comms:
  live: 'Low Earth orbit: the flight surgeon is alerted at the start and sees readings in near real time.'
  loss_of_signal:
  - >-
    The robot pauses any movement and tells the CMO; on the CMO's confirmation it keeps recording on schedule
    and queues everything.
  - >-
    The CMO's sepsis checklist and crew oxygen continue under the onboard checklist and never wait for
    the link or the robot.
  - >-
    The CMO decides under the pre-approved onboard checklist; the robot never waits for the ground before
    urgent onboard care.
  - 'At the next contact the robot sends one package: readings, answers, alerts and decisions.'
  mars_delay:
  - >-
    With up to about 21-23 minutes one way, and blackouts of about 2-3 weeks, the CMO acts on red flags
    onboard.
  - >-
    Alert levels (set against each crewmember's in-flight baseline), the repeat schedule and the onboard
    treatment plan must be set before flight.
  max_delay_for_live_guidance: <set by validation>
sources:
- key: ng253
  text: >-
    National Institute for Health and Care Excellence. Suspected sepsis in people aged 16 or over: recognition,
    assessment and early management. NICE guideline NG253 (2025, last updated Sept 23, 2026).
  url: https://www.nice.org.uk/guidance/ng253/chapter/Face-to-face-assessment
  supports: >-
    Red flags are modeled on NG253 Table 1 (community settings), not the hospital NEWS2 score; the robot
    calculates no score. Do not rely on fever or hypothermia alone (1.4.1); a rise in temperature can
    be a physiological response (1.4.3); heart rate is lower in fit adults and a new arrhythmia can be
    a response to infection; interpret blood pressure against previous values (Table 1: systolic more
    than 40 mmHg below normal); changes in mental state are significant and may be subtle (1.4.6-1.4.7);
    if oxygen saturation is difficult to measure, this may indicate poor peripheral circulation because
    of shock (1.4.10), and pulse oximeters may overestimate saturation, which has been reported in people
    with dark skin; trauma, surgery or invasive procedures in the last 6 weeks is a risk factor; ask how
    often urine was passed in the past 18 hours and about rigors (1.3.5), and about previous presentations
    (1.3.6). Numbers from the guideline are not used as thresholds here.
- key: stahn
  text: >-
    Stahn AC, Werner A, Opatz O, et al. Increased core body temperature in astronauts during long-duration
    space missions. Sci Rep. 2017;7(1):16180.
  url: https://doi.org/10.1038/s41598-017-15560-w
  doi: 10.1038/s41598-017-15560-w
  pmid: '29170507'
  supports: >-
    During exercise, core body temperature rose higher and faster in space than on Earth; resting core
    temperature rose by about 1 °C, developing gradually over 2.5 months of flight.
- key: crucian
  text: >-
    Crucian B, Babiak-Vazquez A, Johnston S, et al. Incidence of clinical symptoms during long-duration
    orbital spaceflight. Int J Gen Med. 2016;9:383-391.
  url: https://doi.org/10.2147/IJGM.S114188
  doi: 10.2147/IJGM.S114188
  pmid: '27843335'
  supports: >-
    Among 46 long-duration ISS crewmembers, upper respiratory symptoms (0.97 per flight year) and various
    other infections were among the most reported events in the categories studied (infections, allergies,
    rashes and hypersensitivities).
- key: ermed
  text: >-
    NASA Human Research Program. Evidence Report: Risk of Adverse Health Outcomes and Decrements in Performance
    due to In-Flight Medical Conditions (2017).
  url: https://humanresearchroadmap.nasa.gov/evidence/reports/Medical.pdf
  supports: >-
    Table 2: 33 respiratory infections, 13 skin infections and 9 urinary tract infections among US astronauts
    (Shuttle, Mir and ISS through 2006).
- key: eau
  text: European Association of Urology. EAU Guidelines on Urolithiasis.
  url: https://uroweb.org/guidelines/urolithiasis
  supports: >-
    Section 3.4.2: the obstructed kidney with all signs of urinary tract infection and/or anuria is a
    urological emergency; urgent decompression is often necessary, with antibiotics.
- key: nice150
  text: >-
    National Institute for Health and Care Excellence. Headaches in over 12s: diagnosis and management.
    Clinical guideline CG150.
  url: https://www.nice.org.uk/guidance/cg150/chapter/Recommendations
  supports: >-
    Headache features that need further investigation or referral include worsening headache with fever,
    sudden-onset headache reaching maximum intensity within 5 minutes, new-onset neurological deficit,
    and impaired level of consciousness.
- key: tb044
  text: NASA OCHMO. Medical Operations. Technical Brief OCHMO-TB-044 (2025).
  url: https://www.nasa.gov/wp-content/uploads/2025/08/ochmo-tb-044-medical-operations.pdf
  supports: >-
    The crew surgeon keeps medical authority when care is delegated; crew medical officers train on the
    JSC 48522 medical checklist.
- key: v1
  text: 'NASA. NASA-STD-3001, Space Flight Human-System Standard, Volume 1: Crew Health, Revision C (2023).'
  url: https://www.nasa.gov/wp-content/uploads/2023/11/nasa-std-3001-vol-1-rev-c-with-signature.pdf
  supports: Requirement IDs beginning V1.
- key: v2
  text: >-
    NASA. NASA-STD-3001, Volume 2: Human Factors, Habitability, and Environmental Health (Rev F, 2026),
    including Section 10.6 Automated and Robotic Systems.
  url: https://standards.nasa.gov/standard/NASA/NASA-STD-3001_VOL_2
  supports: Requirement IDs beginning V2.
- key: marsdelay
  text: >-
    NASA. Mars communications disruption and delay (fact sheet, 2023 Moon to Mars Architecture Concept
    Review).
  url: https://www.nasa.gov/wp-content/uploads/2024/01/mars-communications-disruption-and-delay.pdf
  supports: >-
    One-way Earth-Mars delay up to about 21-23 minutes, plus blackouts of about 2-3 weeks near solar conjunction.
- key: pccp
  text: >-
    U.S. Food and Drug Administration. Marketing submission recommendations for a predetermined change
    control plan for artificial intelligence-enabled device software functions (guidance).
  url: >-
    https://www.fda.gov/regulatory-information/search-fda-guidance-documents/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence
  supports: A model for planning, validating and rolling back software updates.
- key: faa460
  text: 'U.S. Code of Federal Regulations, Title 14, Part 460: Human space flight requirements.'
  url: https://www.ecfr.gov/current/title-14/chapter-III/subchapter-C/part-460
  supports: >-
    Federal rules for crew and spaceflight participants on licensed commercial flights; they set no medical
    standards for spaceflight participants.
corrections:
- date: '2026-10-20'
  version: '0.2'
  source: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 20, 2026 (batch 2 review)
  items:
  - >-
    Red flags are described as modeled on NG253 Table 1 (community settings), not NEWS2; the robot calculates
    no score; rigors are an assessment question, escalated by this draft (item 23).
  - >-
    Low blood pressure is also judged against the in-flight baseline; added in-flight baseline heart rate
    and blood pressure, and RF13 for a new irregular heartbeat flagged by the monitor (item 24).
  - >-
    An oximeter that cannot get a reading now triggers an URGENT alert (possible poor circulation from
    shock), and the probe site and device are recorded (item 25).
  - >-
    A question the patient does not answer (rather than declines) is now treated as possible new confusion:
    the CMO checks alertness at once (item 26).
  - >-
    Added questions about any injury, procedure or catheter in the past 6 weeks and earlier reports of
    feeling unwell (item 27).
  - RF2 (low temperature) now cites NG253 (item 28).
  - >-
    RF10 notes that drainage and antibiotics may not be available onboard and the flight surgeon decides
    on evacuation [V1 3007]; cites EAU section 3.4.2 (item 29).
  - >-
    The CMO is prompted at S1 to decide masks, gloves and distancing before others approach; masks are
    brought first (item 30; no guideline text verified for this order).
  - >-
    Each red-flag decision now applies the action listed for each matching red flag (URGENT alert, stop,
    or private flag), not URGENT for everything (review item 47).
  - >-
    Repeat check-ins inside an open log are now on-screen prompts; the robot approaches only when the
    crewmember or CMO accepts, and a prompt never starts it (item 49, [V2 10174]).
  - >-
    Red flags that had no source now cite one where a source fits; the rest are shown as general clinical
    red flags with no source cited (item 50).
  - >-
    The "item that failed a check" lockout is kept and now notes that it applies when RMP-PHARM-001 is
    used (item 51).
  - NASA evidence reports are cited as "approved for public release" (item 48).
- date: '2026-10-21'
  version: '0.3'
  source: Batch 7 editorial (AI-assisted), Oct 21, 2026
  items:
  - >-
    Citation labels changed to the site's year-only convention: "(approved for public release <date>)"
    now reads "(YYYY)", and NASA-STD-3001 Volume 2 "Revision F (approved July 14, 2026)" now reads "(Rev
    F, 2026)" (sources: ermed, v2). Editorial; no change to steps, thresholds or limits.
- date: '2026-10-21'
  version: '0.4'
  source: Batch 7 review and batch 8 editorial (AI-assisted), Oct 21, 2026
  items:
  - >-
    X5 and the comms loss-of-signal branch now say only the robot pauses: the CMO's sepsis checklist and
    crew oxygen continue under the onboard checklist and never wait for the link or the robot (batch 7
    review item 13). Editorial; robot behaviour unchanged.