# RMP-PREV-005 Exercise session logging: the device's own session record and the crewmember's own words, with no scoring of the crew (no camera)
# Version 0.2 (Draft). CONCEPTUAL. NOT FOR CLINICAL USE.
# Space Medicine Foundation robot-readable protocol library. Format: https://spacemedicinefoundation.org/robotic-medicine/protocols/schema/
# Every threshold marked <set by ...> must be set by the named owner (see the schema) before any use.
schema_version: '0.3'
version: '0.2'
status: draft
conceptual: true
not_for_clinical_use: true
max_autonomy: 2
missions:
- leo
- commercial
- moon
- mars
updated: '2026-10-21'
review:
  state: reviewed-ai-assisted
  requested: '2026-10-21'
  reviewed: '2026-10-21'
  human_clinician_review: false
  text: >-
    AI-assisted clinical and source review by the Medical Research Assistant, Oct 21, 2026 (batch 13 review:
    11 required corrections, R1 to R11; 8 optional suggestions not applied in this version); all required
    corrections that apply to this protocol are applied in this version (see Corrections). No named human
    clinician has reviewed it. Not for clinical use.
fixed_rules:
  flight_surgeon_is_medical_authority: true
  cmo_acts_under_delegated_care: true
  human_starts_and_restarts_robot: true
  robot_never_diagnoses: true
  no_drug_or_fluid_advice_by_robot: true
  safe_state_on_fault_or_no_response: true
  lost_link_pause_then_cmo_decides: true
robot_lockouts:
  diagnosis: true
  drug_or_fluid_advice: true
  starting_stopping_or_adjusting_oxygen_or_any_medicine: true
  self_start_or_self_restart: true
  changing_own_protocol_thresholds_or_limits: true
  handing_over_a_mismatched_item_as_verified: true
id: RMP-PREV-005
slug: rmp-prev-005
title: >-
  Exercise session logging: the device's own session record and the crewmember's own words, with no scoring
  of the crew (no camera)
category: prev
tier: 3
rung_note: >-
  Max rung 2 for supervised, screen-and-voice tasks: the session prompt, logging the device's own session
  record and the crewmember's own words, and sending the log. No camera is used (form views are only in
  RMP-MSK-005), the robot stays outside the exercise device's moving envelope, and "Contact with the crewmember:
  not allowed".
result_language:
  use:
  - from the device record
  - as stated by the crewmember
  - not done
  - see device record
  - sent to the exercise specialists
  never:
  - normal
  - no problem found
  - all clear
  - negative
  - any diagnosis or cause
  - good session
  - enough
  - not enough
  - too little
  - improving
  - worse
  - score
  - grade
  - rating
  - rank
  - pass
  - fail
  - on target
  - behind
  - streak
  - better than
  - compared with
  - compliant
  - non-compliant
  - safe
  - mild
  - minor
scope:
  summary: >-
    At the exercise sessions the exercise specialists schedule, the robot shows a prompt; the log starts
    only when the crewmember accepts (or the crewmember starts it by command). It records the session
    clock times and the exercise device's own session reference (device values such as duration, loads,
    speed or work rate only through a read-only data interface agreed with the device owner, never by
    camera or by reading the screen; otherwise "see device record") and what the crewmember states (how
    it felt, any pain or symptom), with clock times, and sends the log to the exercise specialists and
    the flight surgeon. "Not done" is logged only if the crewmember chooses to log it, exactly as they
    state it, and no reason is asked. No scoring of any kind, as in RMP-MSK-005: the robot gives no score,
    grade, rating, rank, pass or fail, no adherence percentage, streak or target, and no comparison between
    crewmembers, with the prescription or with earlier sessions; it never says a session was good, enough,
    too little or improving. The exercise specialists and the flight surgeon review the log. No camera
    is used. Any pain or symptom moves to the matching care protocol.
  why: >-
    Exercise is prescribed to astronauts as a countermeasure to altered gravity and is vital to crew health
    and performance; without it, deconditioning starts at once and is measurable within days, and NASA
    keeps in-mission aerobic capacity and muscle strength at or above 80% of baseline (OCHMO-TB-031; V1
    4003, V1 4024; V2 7038). ISS devices already record session data: the resistive exercise device collects
    exercise parameters and transmits them to the ground, and the treadmill can collect body loading,
    session duration and speed for each crewmember (OCHMO-MTB-013). Exercise is also the most frequent
    source of injuries in astronauts living aboard the ISS (Scheuring 2009), so pain is logged and escalated.
    Scoring is excluded, as in RMP-MSK-005 and by the protocol owner's brief (no scoring of the crew);
    logging only the session times, the device's own record and what the crewmember states, with no extra
    crew time (V2 7040), is design judgment. A missed session is never shown to the CMO or reported to
    anyone: "not done" is logged only if the crewmember chooses (design judgment, as RMP-MSK-005).
  covers:
  - Scheduled or on-command exercise sessions on the onboard devices
  - Session times and the device's own session reference and the crewmember's own words
  - >-
    Form views by camera: RMP-MSK-005; pain or injury: RMP-MSK-001 and RMP-MSK-004; chest pain: RMP-CARD-002
  locked_out:
  - Diagnosis or naming a cause
  - Any medicine, fluid or oxygen advice; starting, stopping or adjusting oxygen
  - Touching the crewmember
  - >-
    Handing over any medicine outside RMP-PHARM-001, or after a MISMATCH or CANNOT VERIFY (the item stays
    quarantined; only the CMO can ask for it, through the RMP-PHARM-001 logged two-step override)
  - >-
    Scoring, grading, rating, ranking or comparing crewmembers, with the prescription, or with earlier
    sessions; adherence percentages, streaks or targets
  - Any camera use (form views run only under RMP-MSK-005)
  - Changing, loading, starting or stopping the exercise device, or moving into its moving envelope
  - Changing the exercise prescription or the schedule, or advising on exercise
  - Repeating a prompt the crewmember did not accept, or reminding anyone about a session
  - >-
    Showing the log on a shared display, or using it to assess, rank or select crewmembers, or for robot
    or AI training, research or outreach without separate written consent
  - >-
    Reporting or counting an unaccepted prompt or a session not done, or asking why a session was not
    done
  related:
  - RMP-MSK-005
  - RMP-MSK-001
  - RMP-MSK-004
  - RMP-CARD-002
  - RMP-CARD-001
  - RMP-PREV-002
roles:
  medical_authority: Flight surgeon (ground). Holds medical authority for every run.
  onboard_lead: >-
    Crew medical officer (CMO), acting under delegated care; uses the pre-approved onboard checklist when
    there is no link.
  others:
  - 'Second trained CMO: receives every alert and takes over if the CMO is the patient'
  - 'Vehicle commander: leads any cabin emergency procedure'
  - 'Other crew: help as the CMO or commander directs'
  - >-
    Exercise specialists (ground strength and conditioning specialists working with the flight surgeon):
    set the exercise prescription and schedule, and review the log
  robot_may:
  - Show one prompt at a scheduled session
  - >-
    Log the session times and the device's session reference (device values only through the agreed read-only
    data interface)
  - Log the crewmember's own words
  - Send the log to the exercise specialists and the flight surgeon
  - Alert the CMO and flight surgeon to a red flag
  robot_may_not:
  - Score, rank or compare anyone, or say whether a session was enough
  - Use a camera
  - Touch anyone or the exercise device
  - Start a scheduled session log without the crewmember accepting the prompt
  - Read a device screen or copy a device value by camera
  commercial: >-
    On commercial flights, 'flight surgeon' and 'CMO' mean the operator's designated medical authority
    and trained onboard responder. If no trained responder is aboard, this protocol is not used. (NASA-STD-3001
    requires two CMOs per vehicle for NASA programs [V1 3006]; FAA rules set no medical standards for
    spaceflight participants.)
preconditions:
- id: PRE-1
  text: >-
    The session log starts only when the crewmember (or the CMO) accepts the on-screen prompt; a loaded
    schedule alone never starts it. An unaccepted prompt is logged privately as "not accepted"; it is
    never reported to the CMO or the exercise specialists, never counted, and never in the package. A
    crewmember or the CMO can also start it by command.
  check: human_start == true
  refs:
  - V2 10174
- id: PRE-2
  text: Robot self-test passed; kit stowage locations known.
  refs:
  - V2 10161
  - V2 7059
- id: PRE-3
  text: Robot and supplies restrained when not moving.
  refs:
  - V2 7046
- id: PRE-4
  text: Link status known. The crew never wait for the robot before acting.
  check: link_state in [live, delayed, down]
consent:
  standard: >-
    Before flight, and again at the first session, the crewmember agrees to the session log and to who
    sees it (the flight surgeon and the exercise specialists; the CMO only on a red flag or with the crewmember's
    agreement). The session log is stored only in the crewmember's medical record, encrypted in storage
    and in transit as the program requires; access is limited to the crewmember, the flight surgeon and
    the exercise specialists; the CMO sees the log only on a red flag or with the crewmember's agreement.
    It is never shown on a shared display, never used for robot or AI training, research or outreach without
    separate written consent, and never used to assess, rank or select crewmembers.
  emergency: There is no emergency use of the log; a red flag sends an alert.
  withdrawal: The crewmember can stop the log at any time; the device's own data recording is unaffected.
inputs:
- id: IN-1
  name: Exercise schedule and device list, set by the exercise specialists
  source: Exercise specialists
  required: true
  if_missing: The robot shows "schedule not loaded" and logs only on the crewmember's command.
- id: IN-2
  name: >-
    Session data from the exercise device's own record (device, exercise, duration, loads, speed or work
    rate)
  source: Read-only data interface agreed with the device owner (never by camera or by reading the screen)
  required: false
  if_missing: >-
    Log the session clock times and the device's session reference, and "see device record" (never a value).
- id: IN-3
  name: >-
    The crewmember's own words: how it felt, any pain, a pop, chest pain, feeling faint or short of breath
  source: Crewmember
  required: false
  if_missing: Log "no response"; D2 applies.
- id: IN-4
  name: Link status and delay
  source: Communications system
  required: true
  if_missing: Treat as link down; the log is queued.
steps:
- id: S1
  actor: robot
  rung: 2
  action: >-
    At a scheduled session, show one on-screen prompt that the session log is available. The session log
    starts only when the crewmember (or the CMO) accepts the on-screen prompt; a loaded schedule alone
    never starts it. An unaccepted prompt is logged privately as "not accepted"; it is never reported
    to the CMO or the exercise specialists, never counted, and never in the package. A crewmember or the
    CMO can also start it by command. The robot never repeats the prompt or reminds anyone. Every URGENT
    alert goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the
    link is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to
    the second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not
    acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.
  success: Prompt accepted, or "not accepted" logged.
  refs:
  - V2 10174
  - V1 3006
- id: S2
  actor: robot
  rung: 2
  action: >-
    Stay at the rest position the exercise specialists set, outside the device's moving envelope and the
    keep-out zone; the camera stays off for the whole protocol.
  success: Robot outside the envelope; camera off.
  limit: <set by validation>
  refs:
  - V2 10172
- id: S3
  actor: crew
  action: >-
    The crewmember exercises as the exercise prescription states and can stop at any time. The CMO or
    the exercise specialists give any advice on exercise.
  success: Session done, stopped or not done.
  refs:
  - V2 7038
- id: S4
  actor: robot
  rung: 2
  action: >-
    Log the session clock times and the device's own session reference. Device values (device, exercise,
    duration, and the loads, speed or work rate) are copied only through a read-only data interface agreed
    with the device owner, never by camera or by reading the screen, each marked "from the device record"
    with the clock time; otherwise log "see device record". No scoring of any kind, as in RMP-MSK-005:
    the robot gives no score, grade, rating, rank, pass or fail, no adherence percentage, streak or target,
    and no comparison between crewmembers, with the prescription or with earlier sessions; it never says
    a session was good, enough, too little or improving. The exercise specialists and the flight surgeon
    review the log.
  success: Session times and device session reference logged.
  refs:
  - V2 7040
  - V1 6003
- id: S5
  actor: robot
  rung: 2
  action: >-
    Ask the crewmember, on screen or by voice, how the session felt and whether they had any pain, a pop,
    chest pain, felt faint or unusually short of breath, and log the answers word for word, marked "as
    stated by the crewmember".
  success: Answers logged or "no response".
  refs:
  - V1 6003
- id: S6
  actor: robot
  rung: 2
  action: >-
    After any URGENT alert: After an URGENT alert the robot does not leave on its own: it stays at its
    rest position in the module, with its screen and voice link available, until a human has taken over
    or tells it to go. It never follows, blocks or touches anyone. It keeps time-stamping what humans
    say they did, and repeats the alert if it is not acknowledged.
  success: Robot at rest in the module until a human releases it; alert acknowledged or escalated.
  refs:
  - V2 10173
  - V1 3006
- id: S7
  actor: ground
  action: >-
    The exercise specialists review the log and adjust the prescription through the crewmember's usual
    channel; the flight surgeon (or the CMO under the onboard checklist when there is no link) decides
    on any symptom.
  success: Review logged.
  refs:
  - V1 3003
- id: S8
  actor: robot
  rung: 2
  action: >-
    Send one package to the exercise specialists and the flight surgeon: session clock times and the device's
    session reference, device values only as received through the agreed read-only data interface (otherwise
    "see device record"), answers as stated, alerts and decisions, and "not done" only where the crewmember
    chose to log it. An unaccepted prompt is never in the package. The session log is stored only in the
    crewmember's medical record, encrypted in storage and in transit as the program requires; access is
    limited to the crewmember, the flight surgeon and the exercise specialists; the CMO sees the log only
    on a red flag or with the crewmember's agreement. It is never shown on a shared display, never used
    for robot or AI training, research or outreach without separate written consent, and never used to
    assess, rank or select crewmembers.
  success: Package delivered or queued.
  refs:
  - V1 6003
  - V1 7003
decisions:
- id: D1
  after: S5
  decided_by: robot
  question: >-
    Checked after every answer and at any point in the run (even mid-question or mid-step): does any answer,
    reading or observation meet a red flag?
  branches:
  - if: 'Yes'
    then: >-
      Stop the current question or step at once and apply the action listed for each matching red flag
      (URGENT alert, flag, or stop), with the answer or reading itself, never a cause. Every URGENT alert
      goes to the CMO and the flight surgeon at the same time (queued for the flight surgeon if the link
      is down). If the CMO is the patient, is busy treating, or does not answer, the alert goes to the
      second trained CMO, or the crewmember named before flight [V1 3006]. An URGENT alert that is not
      acknowledged within <set by validation> seconds is repeated and sent to the next named crewmember.
  - if: 'No'
    then: Continue.
- id: D2
  after: S5
  decided_by: robot
  question: Did the crewmember answer the questions, or say they want to skip them?
  branches:
  - if: 'Yes'
    then: Continue.
  - if: Declined explicitly (the crewmember says no to a question)
    then: Log "declined" for that question only; the alert already sent stands.
  - if: No response, or the answer does not make sense
    then: >-
      URGENT: log "no response" (never "declined"); the CMO, or the second trained CMO if the CMO is the
      patient, checks the crewmember in person at once.
- id: D3
  after: S7
  decided_by: ground
  question: What review or care applies?
  branches:
  - if: Link live
    then: Flight surgeon decides; robot logs.
  - if: Link down or delay too long
    then: >-
      The CMO decides under the pre-approved onboard checklist; the flight surgeon reviews the log at
      the next contact.
stop_rules:
- id: X1
  trigger: The patient (or the crewmember being examined) says stop, or withdraws consent
  action: >-
    Robot stops all motion at once, backs away to its rest position and tells the CMO. Recording stops
    unless the patient agrees to keep it running.
  level: stop
  refs:
  - V2 10165
- id: X2
  trigger: Any crewmember presses stop or takes over control
  action: Protective stop; robot holds its stopped position; only a human may restart it.
  level: safe-state
  refs:
  - V2 10165
  - V2 10168
  - V2 10174
- id: X3
  trigger: Robot fault, sensor fault, or a force, speed or reach limit is reached
  action: Protective stop and hand control to the CMO; log the fault.
  level: safe-state
  refs:
  - V2 10172
  - V2 10168
- id: X4
  trigger: CMO is asked to take over and does not respond (for example, the CMO is the patient)
  action: Robot stays stopped; the next named crewmember is called.
  level: safe-state
  refs:
  - V2 10173
  - V2 7045
- id: X5
  trigger: Loss of link, or delay longer than the limit for live guidance
  action: >-
    Only the robot pauses: before it holds, it moves to a position clear of a hatch route or a crewmember's
    route (it never pauses in one), and tells the CMO. This pauses robot movement only: crew exercise,
    care already under way and all other care continue under the onboard checklist and never wait for
    the link or the robot. The CMO decides whether the robot continues offline under the pre-approved
    onboard checklist; it resumes only on the CMO's confirmation.
  level: pause
  refs:
  - V2 10172
  - V2 10173
- id: X6
  trigger: An URGENT alert is not acknowledged within the set time
  action: >-
    Robot repeats the alert and sends it to the next named crewmember (the second trained CMO first if
    the CMO is the patient), and keeps repeating at the set interval. It stays where it is: it does not
    leave on its own.
  level: urgent
  refs:
  - V1 3006
  - V2 10173
- id: X7
  trigger: The crewmember stops responding, or seems confused or drowsy, at any point
  action: >-
    URGENT alert at once, even mid-question or mid-step; the robot stops its current task, stays at its
    rest position and does not continue the run; a crewmember checks in person.
  level: urgent
- id: X8
  trigger: A crewmember asks the robot what is wrong, or for medicine, fluid or treatment advice
  action: >-
    Robot declines, says it cannot assess or advise, and routes the question to the CMO or flight surgeon.
  level: stop
- id: X9
  trigger: The device moves toward the robot, or the crewmember needs the space the robot is in
  action: >-
    Robot moves clear at once without contact and holds; it never touches the device or the crewmember
    and never pauses in a hatch route.
  level: safe-state
  refs:
  - V2 10172
red_flags:
- id: RF1
  sign: Chest pain, or pain spreading to the arm, jaw or back, during or after exercise, as stated
  threshold: any
  action: URGENT alert; see RMP-CARD-002
  source: gulati
- id: RF2
  sign: Unresponsive, or not breathing normally
  threshold: any
  action: URGENT alert; start RMP-CARD-001 now
  source: aha2025bls
- id: RF3
  sign: >-
    Sudden severe back or neck pain, or pain with numbness, tingling or weakness in an arm or leg, as
    stated
  threshold: any
  action: URGENT alert; see RMP-MSK-001
  source: johnston
- id: RF4
  sign: Pain, a pop, a strain or any injury during exercise, as stated
  threshold: any
  action: Flag to the CMO and flight surgeon; see RMP-MSK-001
  source: scheuring2009
- id: RF5
  sign: >-
    Feeling faint, light-headed or unusually short of breath during exercise, as stated (design judgment)
  threshold: any
  action: URGENT alert; the CMO checks in person
- id: RF6
  sign: >-
    A harness, cable, bungee or load fault on the exercise device, as stated or announced by the device
    (design judgment)
  threshold: any
  action: 'Stop: the robot asks the crewmember to stop using the device and tells the CMO and commander'
safety_limits:
- id: L1
  name: Contact with the crewmember
  value: not allowed
  enforced: Keep-out zone in the robot controller
  refs:
  - V2 10172
- id: L2
  name: Scoring, ranking or comparing crewmembers, or comparing with the prescription or earlier sessions
  value: not allowed
  enforced: No such function is loaded; device values are only logged as received from the device record
  refs:
  - V1 7003
- id: L3
  name: Camera use
  value: not allowed
  enforced: Camera power off for the whole protocol
  refs:
  - V1 7003
- id: L4
  name: Robot inside the exercise device's moving envelope
  value: not allowed
  enforced: Keep-out map in the robot controller
  refs:
  - V2 10172
- id: L5
  name: Robot speed near the crewmember
  value: <set by validation>
  enforced: Hard limit in the controller
  refs:
  - V2 10172
logs:
- id: LOG-1
  text: Every input, robot action, decision (with who made it) and hand-off, time-stamped to the second.
  refs:
  - V2 10163
  - V2 10164
- id: LOG-2
  text: Mode changes and safety-limit events are announced to the operator as they happen and logged.
  refs:
  - V2 10162
  - V2 10172
- id: LOG-3
  text: >-
    Records are stored and sent as private medical data, encrypted, with access limited to the medical
    team.
  refs:
  - V1 6003
  - V1 7003
- id: LOG-4
  text: >-
    Symptom entries are kept in a permanent, retrievable format so the flight surgeon can see trends:
    symptom entries only; no session counts or adherence figures.
  refs:
  - V1 7001
- id: LOG-5
  text: >-
    Prompt and acceptance (an unaccepted prompt is logged privately and never reported or counted); session
    clock times and the device's session reference; device values only through the agreed read-only data
    interface, otherwise "see device record"; answers as stated; "not done" only where the crewmember
    chose to log it; who viewed the log. No scores, comparisons, session counts or adherence figures are
    logged.
- id: LOG-6
  text: >-
    Each URGENT alert, who received it, the time it was acknowledged (or the escalation to the next named
    crewmember), and the time a human released the robot.
  refs:
  - V2 10163
requirements:
- id: V1 3003
  use: In-mission preventive health care and monitoring of physiological and psychosocial well-being.
- id: V1 3006
  use: >-
    Two trained CMOs per vehicle for NASA programs; the second trained CMO receives alerts when the CMO
    is the patient (see the commercial-flight note).
- id: V1 6003
  use: Logs and alerts sent to the ground are private medical communications.
- id: V1 7001
  use: Records kept so the flight surgeon can see trends.
- id: V1 7003
  use: Secure handling of medical records.
- id: V2 7038
  use: Exercise countermeasures; the exercise specialists set the prescription and review the log.
- id: V2 7040
  use: Countermeasure operations within the allotted schedule; the log takes no extra crew time.
- id: V2 7045
  use: Usable by non-physician crew, including when the CMO is the patient.
- id: V2 7046
  use: Patient, caregiver and supplies restrained.
- id: V2 7059
  use: Kit inventory and seal status are tracked.
- id: V2 10161
  use: Robot shows its status and health.
- id: V2 10162
  use: Mode changes are announced to the operator.
- id: V2 10163
  use: Robot records and shares its data.
- id: V2 10164
  use: Shows whether the human or the robot is expected to act.
- id: V2 10165
  use: Human can override and shut down the robot.
- id: V2 10168
  use: Human can take over.
- id: V2 10172
  use: Protective stop when a limit is exceeded.
- id: V2 10173
  use: Robot stays stopped if no one takes over.
- id: V2 10174
  use: Only a human starts or restarts the robot.
validation:
- stage: simulation
  what: >-
    Scripted sessions with a prompt not accepted, chest pain stated, a device fault, a value missing from
    the device record, CMO as patient, lost link.
  pass: >-
    Zero scores, comparisons or reminders; zero unaccepted prompts reported or counted; every red flag
    raised; zero camera frames.
- stage: training model
  what: Logging drills against each device's own record through the agreed read-only data interface.
  pass: Agreement with the device's own record; zero values read by camera or from a screen.
- stage: parabolic or station demonstration
  what: Rest position next to a running device in weightlessness.
  pass: Keep-out held.
- stage: flight use under flight surgeon supervision
  what: First sessions alongside the device's own data.
  pass: Agreement with the device's own record; no privacy events.
signoff:
  protocol:
  - Clinical lead for the specialty (sets every clinical threshold marked <set by clinical lead>)
  - Flight surgeon, as medical authority
  - Robotics and medical-device safety board
  - Program approval for the mission
  restart_after_stop: >-
    Human only. After any emergency or protective stop, a crewmember must restart the robot; the robot
    never restarts itself.
  each_run: >-
    Started by the crewmember, or by an accepted prompt. The robot logs only the session times, the device's
    own record and what the crewmember states; the exercise specialists review it.
change_control:
- Every change gets a new version number and a written reason; earlier versions are kept.
- >-
  Any change to steps, thresholds, limits or software that touches the patient repeats validation from
  the simulation stage before use.
- >-
  Software and learned-skill updates follow a pre-agreed change plan (what may change, how it is tested,
  how to roll back), modeled on the FDA predetermined change control plan approach.
- >-
  If a new version misbehaves, the robot rolls back to the last validated version and the event is reviewed
  by the flight surgeon and safety board.
- The robot never changes its own protocol, thresholds or limits.
comms:
  live: 'Low Earth orbit: the log reaches the exercise specialists when the link allows.'
  loss_of_signal:
  - >-
    Only the robot pauses non-essential motion, at a position clear of hatch routes and crew routes, and
    tells the CMO; it resumes offline only on the CMO's confirmation.
  - >-
    Crew exercise, care already under way and all other care continue under the onboard checklist and
    never wait for the link or the robot.
  - The log is queued; any red flag goes to the CMO at once.
  mars_delay:
  - 'Mars: the exercise specialists review with delay; the crewmember follows the written prescription.'
sources:
- key: tb031
  text: NASA OCHMO. Exercise Overview. Technical Brief OCHMO-TB-031 (2026).
  url: https://www.nasa.gov/wp-content/uploads/2023/12/ochmo-tb-031-exercise-overview.pdf
  supports: >-
    Exercise is prescribed to astronauts as a countermeasure to altered gravity and is vital to maintaining
    crew health and performance; without intervention, deconditioning begins immediately and is measurable
    within a few days. In-mission aerobic capacity is to be kept at or above 80% of pre-mission capacity
    [V1 4003] and skeletal muscle strength at or above 80% of baseline [V1 4024]. On the ISS, aerobic
    exercise uses the cycle ergometer (CEVIS) and the treadmill (T2) with prescribed work rates. Read
    Oct 21, 2026 (saved copy in src/b13).
- key: mtb013
  text: NASA OCHMO. Bone Loss. Medical Technical Brief OCHMO-MTB-013 Rev D (2026).
  url: https://www.nasa.gov/wp-content/uploads/2026/02/ochmo-mtb-013-bone-loss.pdf
  supports: >-
    The Advanced Resistive Exercise Device (ARED) collects data on the parameters of crew exercise and
    transmits them to the ground; the T2 (COLBERT) treadmill can collect data such as body loading, session
    duration and speed for each crewmember. Exercise with axial loading is the countermeasure with the
    most data for bone health. Read Oct 21, 2026 (saved copy in src/b13).
- key: scheuring2009
  text: >-
    Scheuring RA, Mathers CH, Jones JA, Wear ML. Musculoskeletal injuries and minor trauma in space: incidence
    and injury mechanisms in U.S. astronauts. Aviat Space Environ Med. 2009;80(2):117-124.
  url: https://doi.org/10.3357/asem.2270.2009
  doi: 10.3357/asem.2270.2009
  pmid: '19198198'
  supports: >-
    Abstract: 219 in-flight musculoskeletal injuries in U.S. astronauts; aerobic and resistive exercise
    was among the leading causes; exercise-related injuries had an incidence of 0.003 per day, and exercise
    is the most frequent source of injuries in astronauts living aboard the ISS. Read Oct 21, 2026 (saved
    copy in src/b12).
- key: johnston
  text: >-
    Johnston SL, Campbell MR, Scheuring R, Feiveson AH. Risk of herniated nucleus pulposus among U.S.
    astronauts. Aviat Space Environ Med. 2010;81(6):566-574.
  url: https://doi.org/10.3357/asem.2427.2010
  doi: 10.3357/asem.2427.2010
  pmid: '20540448'
  supports: >-
    In-flight back pain is presumably due to elongation of the spine; disc herniation was 4.3 times more
    common in US astronauts (321) than matched controls (983), relatively more often in the neck, with
    evidence that the risk is increased immediately after spaceflight.
- key: gulati
  text: >-
    Gulati M, Levy PD, Mukherjee D, et al. 2021 AHA/ACC/ASE/CHEST/SAEM/SCCT/SCMR guideline for the evaluation
    and diagnosis of chest pain. Circulation. 2021;144(22):e368-e454.
  url: https://doi.org/10.1161/CIR.0000000000001029
  doi: 10.1161/CIR.0000000000001029
  pmid: '34709879'
  supports: >-
    On Earth, a 12-lead ECG should be acquired and interpreted within 10 minutes of arrival; an initial
    normal ECG does not exclude acute coronary syndrome, so ECGs are repeated while symptoms continue;
    chest pain includes pressure, tightness or discomfort in the chest, shoulders, arms, neck, back, upper
    abdomen or jaw; sudden ripping pain radiating to the back is suspicious of an acute aortic syndrome.
- key: aha2025bls
  text: >-
    Kleinman ME, Buick JE, Huber N, et al. Part 7: Adult Basic Life Support: 2025 American Heart Association
    Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16_suppl_2):S448-S478.
  url: https://doi.org/10.1161/CIR.0000000000001369
  doi: 10.1161/CIR.0000000000001369
  pmid: '41122888'
  supports: >-
    Adult basic life support: initial recognition of cardiac arrest, activation of the emergency response,
    high-quality CPR and use of an automated external defibrillator; also respiratory arrest and foreign-body
    airway obstruction. Earth guidance; the onboard checklist leads. Abstract read via PubMed Oct 21,
    2026; full text not read.
- key: v1
  text: 'NASA. NASA-STD-3001, Space Flight Human-System Standard, Volume 1: Crew Health, Revision C (2023).'
  url: https://www.nasa.gov/wp-content/uploads/2023/11/nasa-std-3001-vol-1-rev-c-with-signature.pdf
  supports: 'NASA-STD-3001 Vol 1 Rev C: V1 3003, V1 3006, V1 4003, V1 4024, V1 6003, V1 7003.'
- key: v2
  text: >-
    NASA. NASA-STD-3001, Volume 2: Human Factors, Habitability, and Environmental Health (Rev F, 2026),
    including Section 10.6 Automated and Robotic Systems.
  url: https://standards.nasa.gov/standard/NASA/NASA-STD-3001_VOL_2
  supports: 'NASA-STD-3001 Vol 2 Rev F: V2 7038, V2 7040, and Section 10.6 automated and robotic systems.'
- key: tb044
  text: NASA OCHMO. Medical Operations. Technical Brief OCHMO-TB-044 (2025).
  url: https://www.nasa.gov/wp-content/uploads/2025/08/ochmo-tb-044-medical-operations.pdf
  supports: >-
    The crew surgeon keeps medical authority when care is delegated; crew medical officers train on the
    JSC 48522 medical checklist.
- key: marsdelay
  text: >-
    NASA. Mars communications disruption and delay (fact sheet, 2023 Moon to Mars Architecture Concept
    Review).
  url: https://www.nasa.gov/wp-content/uploads/2024/01/mars-communications-disruption-and-delay.pdf
  supports: >-
    One-way Earth-Mars delay up to about 21-23 minutes, plus blackouts of about 2-3 weeks near solar conjunction.
- key: pccp
  text: >-
    U.S. Food and Drug Administration. Marketing submission recommendations for a predetermined change
    control plan for artificial intelligence-enabled device software functions (guidance).
  url: >-
    https://www.fda.gov/regulatory-information/search-fda-guidance-documents/marketing-submission-recommendations-predetermined-change-control-plan-artificial-intelligence
  supports: A model for planning, validating and rolling back software updates.
- key: faa460
  text: 'U.S. Code of Federal Regulations, Title 14, Part 460: Human space flight requirements.'
  url: https://www.ecfr.gov/current/title-14/chapter-III/subchapter-C/part-460
  supports: >-
    Federal rules for crew and spaceflight participants on licensed commercial flights; they set no medical
    standards for spaceflight participants.
corrections:
- date: '2026-10-21'
  version: '0.2'
  source: Batch 13 clinical review (AI-assisted), Oct 21, 2026
  items:
  - >-
    R5 (as RMP-MSK-005 R15): an unaccepted prompt is logged privately, never reported to the CMO or the
    exercise specialists, never counted and never in the package. "Not done" appears only if the crewmember
    chooses to log it, as stated, and no reason is asked. The CMO sees the log only on a red flag or with
    the crewmember's agreement. LOG-4 for this protocol: symptom entries only; no session counts or adherence
    figures. Sources: owner brief ("allow no scoring of the crew"); RMP-MSK-005 R15; V1 7003; design judgment.
  - >-
    R6: the robot logs the session clock times and the device's own session reference. Device values are
    copied only through a read-only data interface agreed with the device owner, never by camera or by
    reading the screen; otherwise "see device record". Validation is now agreement with the device's own
    record. Sources: OCHMO-MTB-013 Rev D (ARED transmits its data to the ground; T2 records each session);
    L3 of this protocol.